Calibration Target Accuracy Verification Using GOM 3D Inspection

Calibration accuracy directly depends on the quality of the calibration target. Even micron-level deviations can introduce significant errors in machine vision systems.
At Calibvision, all calibration targets can be verified using professional 3D optical inspection methods. This page demonstrates how we evaluate geometric accuracy using a GOM 3D inspection system.

GOM 3D Inspection Method

We use a full-field 3D optical measurement system (GOM ATOS) to evaluate the calibration target.
The process includes:
This method enables high-resolution, non-contact measurement of the entire calibration target.

Product Model:VMP-050SG-40-BC
Manufacturer:Calibvision
Inspection System:GOM 3D Optical Measurement System
Report ID:CV-TR-2026-001
Date:2026-03-24

What We Measure

Key Inspection Items: We evaluate the following critical parameters:

Dot spacing accuracy
Feature position accuracy
Global geometric deviation
Surface flatness

Visual Inspection Results

Horizontal Spacing Error

Measured accuracy of dot spacing between column Line 1 2 3 4 5 6 7
Dots between Seg 1-2(μm) 0.3031 0.5636 0.5411 0.4391 0.3116 0.4125 0.4669
Dots between Seg 2-3(μm) 0.1783 0.6655 0.6496 0.3142 0.1718 0.2954 0.3183
Dots between Seg 3-4(μm) 0.2893 0.8104 0.7821 0.1707 0.2895 0.4161 0.1876
Dots between Seg 4-5(μm) 0.4017 0.6753 0.6728 0.3059 0.4244 0.3053 0.6876
Dots between Seg 5-6(μm) 0.9017 0.81 0.8165 0.4079 0.5377 0.4261 0.4075
Dots between Seg 6-7(μm) 0.7933 0.663 0.6991 0.2848 1 0.5476 0.532
Mean 0.4779 0.698 0.6935 0.3204 0.4558 0.4005 0.4333
StdDev 0.271 0.0876 0.09 0.0872 0.2686 0.0844 0.158
Min 0.1783 0.5636 0.5411 0.1707 0.1718 0.2954 0.1876
Max 0.9017 0.81 0.8165 0.4391 0.5377 0.5476 0.6876
VMP-050SG-40-BC Horizontal Spacing Full Consistency Analysis

Vertical Spacing Error

Measured accuracy of dot spacing between column Line 1 2 3 4 5 6 7
Dots between Row 1-2(μm) 0.3746 0.2521 0.3651 0.2558 0.4035 0.5278 0.4081
Dots between Row 2-3(μm) 0.4054 0.5136 0.8696 0.2495 0.3856 0.2544 0.3645
Dots between Row 3-4(μm) 0.586 0.6888 0.825 0.6839 0.5673 0.1713 0.563
Dots between Row 4-5(μm) 0.5522 0.4341 0.2985 0.4222 0.2835 0.1682 0.298
Dots between Row 5-6(μm) 0.5407 0.4157 0.5353 0.4259 0.5303 0.4107 0.2908
Dots between Row 6-7(μm) 0.4405 0.2958 0.1536 0.2921 0.4371 0.0586 0.6749*
Mean 0.4832 0.4333 0.5078 0.3882 0.4346 0.2652 0.4332
StdDev 0.0799 0.1436 0.2652 0.1506 0.094 0.1587 0.1409
Min 0.3746 0.2521 0.1536 0.2495 0.2835 0.1713 0.2908
Max 0.586 0.6888 0.8696 0.6839 0.5673 0.5278 0.563
VMP-050SG-40-BC Vertical Spacing Full Consistency Analysis

Overall dot spacing error in μm

 Direction Samples Mean (μm) StdDev (μm) Max (μm)
Horizontal Segments (X) 42 Segs 0.5315 0.1313 0.9034
Vertical Rows (Y) 42 Gaps 0.4431 0.1415 0.8696
Global Centroid Position 49 Pts 0.4872 0.1255 0.9433

Impact on Calibration Performance

These verified results ensure:
Higher calibration accuracy
Lower reprojection error
Stable long-term performance
Reliable measurement results

Verified Accuracy & Inspection Report Service

At Calibvision, accuracy is not just specified — it can be measured and verified.
Each calibration target can be provided with a dedicated inspection report upon request, using professional GOM 3D optical measurement methods.
Third-party verification can also be arranged upon request, depending on project requirements.